Specifications and Reference Data
November 2017
Scaled Variable - The scaled variable configuration allows
the user to convert a transmitter variable into an alternative
measurement, such as flow, mass, or calibrated level (e.g. 5
point verification).
User Defined Variable - Allows designating more than 200
variables in the device as output variable.
A.2.5 Process temperature and
pressure rating
The following figures give the maximum process
temperature (measured at the lower part of the flange, Tri
Clamp, or threaded connection) and pressure rating for
different antenna types.
Final rating may be lower depending on flange selection.
Figure A-4. Cone Antenna (PTFE Seal)
Pressure
psig (bar)
1450
(100)
725
(50)
562
(38.5)
363
(25)
44 (3)
-15 (-1)
-40
100
-76
(-40)
(38)
(-60)
Code CAC
Code CAB
Code CAA
Code CAD
1.
For antenna type code CAB, at 100 °F (38 °C), the rating decreases with
increasing temperature per ASME B16.5 Table 2-2.2, Class 300.
140
(1)
212
302
392
482
(100)
(150)
(200)
(250)
Figure A-5. Cone Antenna (PEEK Seal)
Pressure
psig (bar)
754
(52)
-22
5
(-30)
(-15)
-15
(-1)
-76
-13
(-60)
(-25)
Code CBF (FVMQ)
Code CBV (Viton)
Code CBM (FKM)
Code CBK (Kalrez 6375)
Figure A-6. Process Seal Antenna
Pressure
psig (bar)
363 (25)
232 (16)
-7 (-0.5)
-76
-13
(-60)
(-25)
Tri Clamp
Temp.
Figure A-7. Parabolic Antenna
)
°F (°C
Pressure
psig (bar)
43 (3)
-7 (-0.5)
-67
(-55)
Reference Manual
00809-0100-4408, Rev BA
Temp.
°F (°C)
338
392
428
482
(170)
(200)
(220)
(250)
Temperature
392
°F (°C)
(200)
Temperature
392
°F (°C)
(200)
Specifications and Reference Data