Configuration
November 2017
5.3
System readiness
5.3.1
Confirm correct device driver
The transmitter meets the NAMUR recommendation NE 53. Verify the latest Device Descriptor (DD) or
FDI Package is loaded on your systems to ensure proper communication.
1. Within
DD or FDI Package.
2. Download the latest DD at EmersonProcess.com/DeviceFiles.
3. Download the latest FDI Package at Emerson.com/RosemountRadarMasterPlus.
Table 5-1. Identification and Compatibility According to NAMUR NE 53
Release
date
NAMUR hardware
(1)
revision
March-17
1.0.xx
1.
NAMUR Revision is located on the transmitter label. Differences in level 3 changes, signified above by xx, represent minor product changes as defined per NE53.
Compatibility and functionality are preserved and product can be used interchangeably.
2.
Device software revision is located on the transmitter label, e.g. 1.A3. It can also be found in Rosemount Radar Master Plus (under Overview, select Device Information
> Revisions).
3.
HART Revision 6 and 7 can be switched in field. Default HART universal revision from factory is located on the transmitter head label, e.g. PROTOCOL 6.
4.
Device revision is located on the transmitter label, e.g. DEVICE REV 1.
5.4
Get started with your preferred configuration tool
The Rosemount
Rosemount Radar Master Plus (running in the Instrument Inspector
Device Descriptor (DD) based systems, e.g. AMS Device Manager, 475 Field Communicator, AMS Trex
Device Communicator, and DeltaV
Field Device Integration (FDI) based systems
Rosemount Radar Master Plus is the recommended tool for configuration.
5.4.1
Rosemount Radar Master Plus
The Rosemount Radar Master Plus is a user-friendly software package that includes basic configuration
options, as well as advanced configuration and service functions. The Instrument Inspector Application
or any FDI compliant host is needed to run Rosemount Radar Master Plus.
Instrument Inspector is shipped with every transmitter. See the CD installation guide for a list of
supported HART modems and system requirements.
Instrument Inspector is also available at:
Emerson.com/InstrumentInspector
68
Table
5-1, use the HART
Device identification
NAMUR software
Device software
(1)
revision
revision
1.0.xx
™
5408 and 5408:SIS Level Transmitters can easily be configured by using:
®
Universal Revision and Device Revision numbers to find the correct
DD and FDI identification
®
HART
universal
(2)
(3)
revision
1.Axx
6
7
™
, or any other EDDL or enhanced-EDDL host
Reference Manual
00809-0100-4408, Rev BA
Review
instructions
Device
Manual document
(4)
revision
number
1
00809-0100-4408
1
™
Application)
Review
functionality
Change
description
N/A
™
Configuration