11.6.4

Span calibration by test chain

Span calibration by test chain
If the scale is loaded with a chain (e.g. 50% of the nominal load), the weight is assigned to
the new digital value returned by the analog-to-digital converter.
Step 1
The belt is warmed-up and the zero point has been determined.
1. Stop the belt.
2. Enter a test chain corresponding to the measuring range in DR3 (e.g. 50% of the nominal
load).
3. Place the test chain onto the scale.
Step 2
1. Switched the belt on. The test chain is positioned on the scale.
2. Start recording of the weight for the parameterized number of belt revolutions using the
command "Start span calibration with test chain" (65). The status display in DR30
indicates that the calibration is running. The result is displayed in DR4.
3. Import the working point into DR3 using the command "Apply span calibration digits" (89).
The characteristic curve has thus been determined. The scale can calculate the weight
values for the complete load range.
11.6.5

Span calibration with known material flow

Span calibration with known material flow
The material test is usually used to determine the correction factor of a calibrated scale.
However, a known quantity of material can also be used to carry out a scale calibration. The
calibration is based on transportation of a known quantity of material, and comprises the
following steps:
Step 1
The belt is warmed-up and the zero point has been determined.
SIWAREX WT241
Manual, 05/2015, A5E36046748Aa
11.6 Calibration of weight measurement