Negative And Positive Zero Offset In; Smallest Totalized Value; Minimum Flow Rate; Maximum Flow Rate - Siemens SIWAREX WT241 Manual

Electronic
Table of Contents

Advertisement

Communication
11.9 DR 6 Limits
11.9.1

Negative and positive zero offset in %

Zeroing sets the current weight of the scale and the belt load to zero. The specification is
made in % of the nominal belt load. The range refers to the original zero point of the last
calibration.
In the case of sales for legal trade operation, the maximum zero offset is 4%.
11.9.2

Smallest totalized value

The smallest totalized value is the smallest value which may be recorded with verification
capability.
11.9.3

Minimum flow rate

Dropping below the minimum flow rate is displayed in the status area of the scale. The
specification is made in % of the design flow rate.
11.9.4

Maximum flow rate

Exceeding the maximum flow rate is displayed in the status area of the scale. The
specification is made in % of the design flow rate.
11.9.5

Delay for flow rate limits

Reaching the maximum flow rate is displayed in the status area of the scale only following
expiry of the delay. The specification is made in ms.
11.9.6

Minimum belt speed

Dropping below the minimum belt speed is displayed in the status area of the scale. The
specification is made in % of the nominal belt speed which was specified during
commissioning or calculated.
11.9.7

Maximum belt speed

Exceeding the maximum belt speed is displayed in the status area of the scale. The
specification is made in % of the nominal belt speed which was specified during
commissioning or calculated.