Chapter 17 Cleaning and Maintenance

Cleaning the Propaq M unit

To clean the Propaq M unit, use a nearly dry cloth containing one of the mild cleaning agents
listed below. DO NOT allow cleaning agent or water to run into the crevices or connector
openings at any time. Thoroughly wipe off any excess cleaning solution from the Propaq M
unit with a dry cloth. Always check monitor and connector opening for unusual wear, damage
or dampness while cleaning
Use only these recommended cleaning agents:
Warm water
Hydrogen peroxide solution
Liquid soap
Wex-cide
Windex
Never use these cleaning agents:
Butyl alcohol
Denatured ethanol
Freon
Mild chlorine bleach solution
Isopropyl alcohol
Trichloroethane, trichloroethylene
Acetone
Vesphene II
Enviroquat
Staphene
Misty
Glutaraldehyde

Cleaning the NIBP Blood Pressure Cuff

Clean the cuff with common hospital disinfectants, including Clorox
isopropyl alcohol, Lysol
gently with the solution, then rinse. DO NOT allow solution to enter cuff tubes. Allow the cuff
and hose to completely dry before patient use.
Cleaning SpO
Sensors
2
Clean the SpO
recommended agents listed above. DO NOT submerge the probe or its connector in any liquids
or cleaning agents. Thoroughly wipe off any excess cleaning solution with a dry cloth. Allow
the sensor to completely dry before patient use.
17-2
®
®
®
solution, Phisohex
sensors with a cloth that has been slightly dampened with one of the
2
www.zoll.com
®
®
®
®
, Quatricide
, Virex
and Vesphene
(1:10 solution),
®
. Wash
9650-000820-01 Rev. K