Cleaning Instructions; Cleaning The Propaq Md Unit; Cleaning The Nibp Blood Pressure Cuff - ZOLL Propaq MD Operator's Manual

Portable defibrillator
Hide thumbs Also See for Propaq MD:
Table of Contents

Advertisement

C
18 M
HAPTER
AINTENANCE

Cleaning instructions

Cleaning the Propaq MD unit

To clean the Propaq MD unit, use a nearly dry cloth containing one of the mild cleaning agents
listed below. DO NOT allow cleaning agent or water to run into the crevices or connector
openings at any time. Thoroughly wipe off any excess cleaning solution from the Propaq MD
unit with a dry cloth.
Always check monitor and connector opening for unusual wear, damage or dampness while
cleaning.
Use only these recommended cleaning agents:
Warm water
Hydrogen peroxide solution
Liquid soap
Wex-cide
Windex
T.B.Q.
Never use these cleaning agents:
Butyl alcohol
Denatured ethanol
Freon
Mild chlorine bleach solution
Isopropyl alcohol
Trichloroethane, trichloroethylene
Acetone
Vesphene II
Enviroquat
Staphene
Misty
Glutaraldehyde

Cleaning the NIBP Blood Pressure Cuff

Clean the cuff with common hospital disinfectants, including Clorox
isopropyl alcohol, Lysol
gently with the solution, then rinse. DO NOT allow solution to enter cuff tubes. Allow the cuff
and hose to completely dry before patient use.
18-8
®
®
®
®
solution, Phisohex
www.zoll.com
®
®
®
®
, Quatricide
, Virex
and Vesphene
(1:10 solution),
®
. Wash
9650-0802-06 Rev. A