ZOLL Propaq M Operator's Manual page 75

Hide thumbs Also See for Propaq M:
Table of Contents

Advertisement

5-Lead Electrode Placement
Depending upon local usage, the ECG leads are marked either RA, LA, LL, RL, and V or R, L,
F, N and C. The following table shows the markings and color codes for the different lead sets.
AHA Color Coding
RA/White Electrode
LA/Black Electrode
LL/Red Electrode
RL/Green* Electrode
V/Brown* Electrode
9650-000820-01 Rev. K
IEC Color Coding
R/Red Electrode
L/Yellow Electrode
F/Green Electrode
N/Black* Electrode
C/White* Electrode
5-Lead Electrode Placement
Propaq M Operator's Guide
ECG Monitoring Setup
Placement of Electrodes
Place near patient's right mid-clavicular line,
directly below clavicle.
Place near patient's left mid-clavicular line,
directly below clavicle.
Place between 6th and 7th intercostal
space on patient's left mid-clavicular line.
Place between 6th and 7th intercostal
space on patient's right mid-clavicular line.
Single movable chest electrode. Place this
electrode in one of the positions, V1 - V6,
as shown in the following figure.
V1 -- 4th intercostal space at right sternal
margin.
V2 -- 4th intercostal space at left sternal
margin.
V3 -- Midway between V2 and V4 leads.
V4 -- 5th intercostal space at mid-clavicular
line.
V5 -- Same transverse level as V4 at left
anterior-axillary line.
V6 -- Same transverse level as V4 at left
mid-axillary line.
6-5