Chapter 7 Monitoring Respiration (Resp) and Heart Rate (HR)

Using the Resp Parameter Control Panel

To configure alarms through the
highlight and select the Respiration Rate meter and display the
The
Resp
• RR/BR Alarm
• No Breath Alarm
"Off".
• Resp Monitoring --
• Resp Lead
Propaq M unit calculates the respiration rate. Resp Lead selection is independent of ECG
Lead selection.
• CO2/Resp Sweep Speed --
Enabling/Disabling Resp Monitoring
Select the Resp Monitoring prompt to enable or disable
Monitoring is set to On (the default), the Propaq M unit displays the Respiration Rate Meter.
When set to off, Propaq M removes the Respiration Rate meter from the display.
Warning!
When using impedance pneumography, the Propaq M unit automatically rejects
cardiovascular artifact (CVA). This function requires the accurate ECG R-wave
detection. Therefore, always select the ECG lead with the most prominent QRS complex
when using impedance pneumography to monitor respiration.
7-4
Resp
Resp
RR/BR Alarm
No Breath Alarm
Resp Monitoring
Resp Lead
CO2/Resp Sweep Speed
Respiration Parameter Control Panel
Parameter Control Panel allows you to set the following parameters:
-- enable/disable the Resp alarm and set high/low alarm limits.
-- sets the duration of the No Breath alarm, or disables alarm by selecting
enable/disable respiration monitoring.
-- selects the Resp lead, Lead I (RA-LA) or Lead II (RA-LL), from which the
sets the respiratory sweep speed on the display.
www.zoll.com
Parameter Control Panel, use the navigation keys to
Lower
Upper
10
30 sec
On
Lead II (RA-LL)
12.5 mm/s
monitoring. When Resp
Resp
Parameter Control Panel:
Resp
30
9650-000820-01 Rev. K