Printer (Recorder)

Type: High-resolution thermal array.
Annotation: Time, date, ECG lead, ECG gain, heart rate and treatment summary events.
Paper Width: 80 mm.
Paper Speed: 25 mm/sec, 50 mm/sec
Delay: 6 seconds.
Frequency Response: Automatically set to monitor's frequency response.
Treatment Summary:
10 switches to record key ACLS events (IV, INTUB, EPI, LIDO, ATROP, etc.).
Automatically logs into memory the type of event, time and ECG sample.
Full Disclosure Case Log: A combination of 32 monitor snapshots; 500 non-ECG events; and
24 hours of continuous ECG (4 waveforms), Capnography and IBP (3 channels). The actual
information stored can be more or less depending on the use profile and the log configuration
settings.
Record Modes: Manual and automatic (User-configurable).

Battery

Type: Rechargeable Lithium-Ion, 11.1Vdc, 6.6 Ah, 73Wh
Capacity:
With a new, fully charged battery operating at room temperature:
At least 7.5 hours of continuous monitoring of ECG, SpO
channels, and 2 channels of Temperature, with NIBP measurements every 15 minutes
(display set to 30%).
Note:
Battery Indicators:
5 Battery capacity LED indicators, Fault indicator, Recalibration indicator
Recharge Rate: 100% in 4 hours, when initiated at Low Battery indication.
9650-000820-01 Rev. K
Proper battery care is required to maintain maximum available capacity.
Propaq M Operator's Guide
Printer (Recorder)
, CO
, three Invasive Pressure
2
2
A-5