Setting Upper And Lower Spoc Alarm Limits; Setting Upper And Lower Pvi Alarm Limits; Setting Upper And Lower Pi Alarm Limits - ZOLL Propaq M Operator's Manual

Hide thumbs Also See for Propaq M:
Table of Contents

Advertisement

Setting Upper and Lower SpOC Alarm Limits

The following table lists the default SpOC limits for adult, pediatric, and neonate patients, and
gives the range in which you can set these limits
Patient Type
Adult
Pediatric
Neonate

Setting Upper and Lower PVI Alarm Limits

The following table lists the default PVI limits for adult, pediatric, and neonate patients, and
gives the range in which you can set these limits
Patient Type
Adult
Pediatric
Neonate

Setting Upper and Lower PI Alarm Limits

The following table lists the default PI limits for adult, pediatric, and neonate patients, and
gives the range in which you can set these limits
Patient Type
Adult
Pediatric
Neonate
9650-000820-01 Rev. K
Enabling/Disabling SpO2 Alarms and Setting Alarm Limits
SpOC Limit Default
Lower: 10 ml/dL
Upper: 25 ml/dL
Lower: 10 ml/dL
Upper: 25 ml/dL
Lower: 10 ml/dL
Upper: 25 ml/dL
PVI Limit Default
Lower: 5%
Upper: 40%
Lower: 5%
Upper: 40%
Lower: 5%
Upper: 40%
PI Limit Default
Lower: 0%
Upper: 20%
Lower: 0%
Upper: 20%
Lower: 0%
Upper: 20%
Propaq M Operator's Guide
SpOC Limit Range
Lower: 0 - 34.9 ml/dL
Upper: 0.1 - 35 ml/dL
Lower: 0 - 34.9 ml/dL
Upper: 0.1 - 35 ml/dL
Lower: 0 - 34.9 ml/dL
Upper: 0.1 - 35 ml/dL
PVI Limit Range
Lower: 0 - 98%
Upper: 2 - 100%
Lower: 0 - 98%
Upper: 2 - 100%
Lower: 0 - 98%
Upper: 2 - 100%
PI Limit Range
Lower: 0 - 19.8%
Upper: 0.2 - 20%
Lower: 0 - 98%
Upper: 2 - 100%
Lower: 0 - 98%
Upper: 2 - 100%
10-13