General - ZOLL Propaq M Operator's Manual

Hide thumbs Also See for Propaq M:
Table of Contents

Advertisement

Specifications

General

Weight:
With Battery: 8.5 lbs.
With Battery and Printer: 9.9 lbs.
Dimensions:
Without Handle: 8.9" x 8.7" x 6.5"
With Handle: 8.9" x 10.4" x 7.1"
With Printer: 8.9" x 10.4" x 7.9"
Operating:
Temperature: 0 to 50° C
Humidity: 15 to 95% RH (non-condensing)
Vibration:
Shock: MIL-STD 810G, Method 516.6, Tested at 75 g
Bump: EN 1789 (IEC 60068-2-29)
Drop: MIL-STD 810G, Method 516-6, Tested at 1 meter with 26 drops
Altitude: -170 M to 4572 M (-557 feet to 15,000 feet)
Transport and Storage:
Temperature: -30 to 70°C
Note:
Humidity: 15 to 95% RH (non-condensing)
Atmospheric pressure: 572 mbar to 1034 mbar
Shock/vibration: MIL STD 810G, Method 514.6, 4.4.2, Procedure II
Safety Classification: Class 1 and internal power per IEC/EN 60601-1
Enclosure Protection:
Solid Foreign Object: IEC 60529, IP5X
Water: IEC 60529, IPX5
Auxiliary Operating Power:
Auxiliary Power Adapter, 8300-0004
Input: 100-240V
Output: 14.5V
IP Rating: IP23
A-6
MIL-STD 810G, Method 514.6, Profile for combined UH-1, UH-60, and CH-47
Rotary Wing Aircraft.
EN 1789 for ambulance
IEC 60601-1, Tested at 2 meters
The Propaq M device may not perform to specifications when stored at the upper or
lower extreme limits of storage temperature and immediately put into use.
50-60 Hz, 2A
100-115V
400 Hz, 2A
4.15A
80W (peak)
www.zoll.com
9650-000820-01 Rev. K

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents