Using The Nibp Parameter Control Panel - ZOLL Propaq M Operator's Manual

Hide thumbs Also See for Propaq M:
Table of Contents

Advertisement

Setting Upper and Lower NIBP MAP Alarm Limits
Initially, the NIBP MAP Alarm Settings menu specifies that NIBP MAP alarms are enabled
(
)
or disabled (
ON
lower limits can be
limits for adult, pediatric, and neonate patients, and gives the range in which you can set these
limits:
Patient Type
Adult
Pediatric
Neonate

Using the NIBP Parameter Control Panel

Use the navigation keys to highlight and select the NIBP numeric display to display the NIBP
Parameter Control Panel:
On the NIBP Parameter Control Panel, you can select an alarm configuration option (
or
) to enable/disable alarms and set high and low alarm limits (as described previously).
MAP
You can also configure these NIBP options:
NIBP Mode (
NIBP Automatic Measurement Interval (
Enable/Disable Smartcuf motion artifact filter
Start/Stop Turbocuf Measurements (
9650-000820-01 Rev. K
),
and displays the default Upper and Lower MAP limits. The upper and
OFF
or
(default is
ON
OFF
MAP Default
Lower: 50 mmHg
Upper: 120 mmHG
Lower: 50 mmHg
Upper: 110 mmHg
Lower: 35 mmHg
Upper: 80 mmHg
NIBP
NIBP Systolic Alarm
NIBP Diastolic Alarm
NIBP MAP Alarm
NIBP Mode
NIBP Auto Mode Interval
SmartCuf On/Off
50
59
NIBP Parameter Control Menu
)
Mode
Propaq M Operator's Guide
Configuring NIBP Alarms and Settings
). The following table lists the default MAP alarm
OFF
MAP Range
Lower: 25-230 mmHg
Upper: 23-228 mmHg
Lower: 25-140 mmHg
Upper: 23-138 mmHg
Lower: 15-110 mmHg
Upper: 13-108 mmHg
Lower
Upper
220
75
35
110
50
120
Manual
5 min
On
Start TurboCuf
150
100
200
)
Auto Interval
(Smartcuf)
)
Turbocuf
250
300
,
SYS
DIA
8-13
,