Chapter 11 Monitoring Invasive Pressures (IBP)

Enabling/Disabling IBP Alarms and Setting Alarm Limits

When enabled, the Propaq M unit sounds alarms whenever IBP measurements are outside set
limits for the following:
High and Low Systolic Pressure
High and Low Diastolic Pressure
High and Low Mean Arterial Pressure (MEAN)
Note:
To enable (or disable) IBP alarms and set Upper and Lower alarm limits and IBP channel, you
can either use the navigation buttons to highlight and select the IBP channel display or:
1. Press the More quick access key (
Press
2.
Press the Limits quick access key (
3.
Use the navigation buttons to highlight and select the appropriate alarm menu selection. For
4.
each IBP channel, there are alarm menus for systolic (Systolic Alarm), diastolic (Diastolic
Alarm) and mean arterial pressure (Mean Alarm) alarm settings. On the selected Alarm
Settings menu, use the navigation keys to select the fields that you want to change. The fields
are
Upper Limit Enable
Lower Limit Enable
Upper Limit
Lower Limit
When you are finished changing values on the alarm menu, navigate to the Backarrow key
5.
to exit the menu.

Setting Upper and Lower Systolic (SYS) Alarm Limits

Initially, the IBP Systolic Alarm Settings menu specifies that the IBP systolic pressure alarms
are enabled (
The upper and lower limits can be
default IBP Systolic limits for adult, pediatric, and neonate patients, and gives the range in
which you can set these limits:
Patient Type
Adult
Pediatric
Neonate
11-6
When enabled, any of the low alarms for IBP (systolic, diastolic, or mean) will also
provide an alert in the event of a disconnected catheter.
.
) or disabled (
ON
OFF
IBP Systolic Limit Default
Lower: 75 mmHg
Upper: 220 mmHg
Lower: 75 mmHg
Upper: 145 mmHg
Lower: 50 mmHg
Upper: 100 mmHg
www.zoll.com
).
).
Limits
)
,
and displays the default upper and lower systolic limits.
or
(default is
ON
OFF
OFF
IBP Systolic Limit Range
Lower: -30 to 298 mmHg
Upper: -28 to 300 mmHg
Lower: -30 to 298 mmHg
Upper: -28 to 300 mmHg
Lower: -30 to 298 mmHg
Upper: -28 to 300 mmHg
). The following table lists the
9650-000820-01 Rev. K