Preparation

Use of accessories

WARNING
Stability may be impaired by the attachment of
additional accessories/loads. In order to ensure
stability (EN 60601-2-21), a one-sided accessory load of
25 Nm* must not be exceeded, when tilted at 10°.
1 Stand column
2 Shelf
3 Base plate
4 Swivel cabinet
5 Compact rail
6 Standard rail

Recommended measures for increasing stability

Attach accessory loads as low and as close to the stand
columns as possible and avoid one-sided loading.
To ensure stability**, observe the following restrictions
when using accessories:
– Do not exceed a maximum patient weight of 8 kg
– Observe a maximum weight limit of 20 kg for the shelf and
a maximum mounting height of 40 cm above the base
plate
– Maximum loading of 2.4 kg for a swivel cabinet,
corresponding to a maximum total weight of 9 kg
– Mount swivel cabinets (2M20638) on opposite sides, if
present.
*
corresponds to approx. 5 kg at a distance of 20 cm from the
stand column, at a height of 2.00 m when tilted at 10°
** EN IEC 60601-2-21 tilt stability ensured up to an inclination of
10°
1
2
5
6
3
4
Preparation
Observe maximum
loading
4
9