Operation
Test Nurse call
P
1 Briefly press
button.
2 The following message appears in the text display:
»Configuration Mode
Mode c06: Nurse Call Relay«.
3 The abbreviation for this mode appears: »Nuc«.
4 The code for the current mode appears: »c06«.
5 Display »0« = Nurse call relay open.
Display »1« = Nurse call relay closed.
G
6 Press
button until the relay switches.
7 Press
button for 4 seconds.
Check
The relay is activated.
The "Inop." alarm is triggered: the red LED »Inop.« lights
up and the continuous alarm sounds.
Select language
P
1 Briefly press
button.
2 The following message appears in the text display:
»Configuration Mode
Mode c07: Display language –> English«.
3 The abbreviation for this mode appears: »LAN«.
4 The code for the current mode appears: »c07«.
5 Indication of the language number.
Factory setting »2« = English.
P
p
6 Press
or
button until the number of the required
language appears on the display.
Possible settings: from 1 to 11.
Set display contrast
P
1 Briefly press
button.
2 The following message appears in the text display:
»Configuration Mode
Mode c08: Display contrast«.
3 The abbreviation for this mode appears: »CON«.
4 The code for the current mode appears: »c08«.
5 Indication of the set contrast. Default setting.
The code for the actual mode appears »128«
P
p
6 Press
or
button until the required contrast has
been set.
Possible settings: 1 to 255 in increments of 1.
46
2
5
10
3
5
4
1
skin C
manual
skin
man .
1
2
5
10
3
5
4
1
skin C
manual
skin
man .
1
6
2
5
10
3
5
4
1
skin C
manual
skin
man .
1
6
Babytherm 8010
6
15 min
C
7
< 36 C
mattress . C
Babytherm 8010
15 min
C
< 36 C
mattress . C
Babytherm 8010
15 min
C
< 36 C
mattress . C