Scale-Tronix ® 4002-Bt; Calibration - Dräger Babytherm 8004 Instructions For Use Manual

Open intensive care, resuscitation
Hide thumbs Also See for Babytherm 8004:
Table of Contents

Advertisement

®
Scale-Tronix
4002-BT
The measurement precision depends on the local geological
and geographical conditions.
The specified precision only applies if the device is calibrated
at the installation site.
Measuring range
Measurement precision (e)
Resolution (d)
Linearity
Electric power supply
Power input
Standards
Dimensions
Weighing unit
Display unit
Weight
Weighing unit
Display unit

Calibration

NOTE
The scale has been calibrated before delivery. In certain
geographical regions, the precision specified in the
Technical Data will only be attained if the device is recali-
brated at the installation site.
Calibration must be performed by authorised personnel
only.
250 g (0.6 lbs) to 8 kg (17.6 lbs)
10 g (0.022 lbs)
10 g (0.022 lbs)
±0.1 % + 1 digit
100 to 127 V AC, 50/60 Hz
220 to 240 V AC, 50/60 Hz
4 W
IEC / EN 60601-1
complies with EN 45 501
67.3 x 33 x 3 cm (26.5 x 13 x 1.2 inches)
43.2 x 36.2 x 6.0 cm (17 x 14.3 x 2.4 inches)
4 kg (8.8 lbs)
3.6 kg (7.9 lbs)
Technical Data
71