Operation
For deviations greater than ±0.5 °C between the set and
measured skin temperature:
1 The skin temperature display starts flashing.
– An audible alarm sounds,
2 the yellow LED starts flashing and
– the red central alarm lamp on the radiant warmer starts to
flash.
3 The following message appears in the text display:
»Skin temp. deviation above 0.5 °C
Press OK button to acknowledge«.
The permitted skin temperature deviation can be set in con-
figuration mode see page 43.
A default skin temperature deviation of ±0.5 °C is set by
the manufacturer.
The audible alarm can be muted for 15 minutes:
G
4 Press
button
or
5 Press OK button.
– The audible alarm is cancelled,
2 the yellow LED lights up,
– the red central alarm lamp on the radiant warmer goes out.
1 Skin temperature display starts to flash.
3 The following message appears in the text display:
»Skin temp. deviation above 0.5 °C«.
After the measured skin temperature has returned to a value
within ±0.5 °C of the set temperature, the yellow LED goes
out, the audible alarm is cancelled and the message in the
display disappears.
Before temporarily removing the skin temperature sensor
from the skin, switch to manual mode (»man.«), see page 23.