Deviation from set temperature
If the deviation between the set and measured mattress tem-
perature is greater than ±1 °C:
– The audible alarm sounds,
1 the yellow LED starts flashing and
– the red central alarm lamp on the radiant warmer flashes.
2 Measured mattress temperature value flashes,
3 The following message appears in the text display:
»Mattress temp. deviation above 1 °C
Press OK button to acknowledge«.
The audible alarm can be muted for 15 minutes:
G
4 Press
button
or
5 Press the OK button
– The audible alarm is cancelled,
1 the yellow LED is lit continuously and
– the red central alarm lamp on the radiant warmer goes out.
3 The following message appears in the text display:
»Mattress temp. deviation above 1 °C«
2 Measured mattress temperature value flashes
When the measured mattress temperature returns to within
±1 °C of the set value:
1 The yellow LED goes out and
– the audible alarm is switched off.
After initially switching on the system:
– the audible alarm is suppressed during the warm-up
phase:
1 The yellow LED is lit.
3
10
5
1
skin C
manual
mattress . C
skin
man .
Operation
Babytherm 8010
1
4
15 min
2
C
< 36 C
5
33