Configuration mode
WARNING
No care/therapy should be carried out during configura-
tion of the system, because the normal equipment func-
tions (e.g. temperature measurement, alarms) are
switched off.
Capabilities available in configuration mode
Code
Activity
c01
Show current software version
c02
Set skin temperature deviation
c03
Set volume of soft alarm
c04
Operating hours of phototherapy
c05
Test control panel buttons
c06
Test Nurse call
c07
Set language
c08
Set display contrast
Err
Error mode
The significance of the keys and displays on the control panel
is different from that applicable when working with the radiant
warmer!
1 Activate/deactivate configuration mode
2 Information on the activated mode
3 Abbreviation for the activated mode
4 Code for the activated mode
5 Set next/previous mode
6 Increase/decrease values for the activated mode
7 Activate/deactivate error mode
8 Change states for the activated mode
9 Show setting in current mode
Activate configuration mode
1 Press OK button for 3 seconds.
A brief acoustic signal sounds.
2 The following message appears in the text display:
»Configuration Mode
Press OK button to start«.
1 Press OK button again. A brief acoustic signal sounds.
Configuration mode is now active. The equipment is in
mode »c01« and shows the current software version.
2
9
10
3
5
4
1
skin C
manual
7
skin
man .
6
5
2
10
5
1
skin C
manual
skin
man .
Operation
Babytherm 8010
15 min
C
8
< 36 C
mattress . C
1
Babytherm 8010
15 min
C
< 36 C
mattress . C
1
43