Shelf
The unit is fitted with one shelf.
Make sure that the shelf is firmly in position.
To fit rails to the shelf:
1 Screw rail to the left and/or right of the shelf.
Place desired auxiliary equipment on the rails and shelf.
●
Max. 20 kg per shelf.
2 Fasten infusion holder to stand column at the required
height.
Compact rail/rail brackets
This rail is designed to hold auxiliary equipment,
e.g. O
flowmeter, O
monitor, infusion pumps.
2
2
3 Fasten the compact rail(s) to one column of the stand at
the required height.
4 Fasten rail brackets and
5 small equipment bar to one column of the stand at the
required height.
Swivel tray
For small articles and parts. Max. load 3 kg.
Position the swivel tray on one of the columns of the stand
●
and tighten the hand screw.
Recommended height: level with the mattress.
Make sure that the swivel path is kept clear of obstructions!
●
Ensuring stability
Stability is ensured when the unit is placed on a surface
with a slope of not more than 10°.
The following restrictions apply when adding accessories:
Do not exceed the maximum load-bearing capacity of
●
the stand = 30 kg.
If accessories are positioned on a column of the stand,
●
e.g. on two swivel trays, the maximum load for this col-
umn is 5 kg.
2
1
4
3
5
3
4