Testing Readiness for Operation
Testing Readiness for Operation
Before using for the first time
1 Check that the mains voltage matches the values indicated
on the rating plate next to the power cable.
2 Plug into the mains.
®
Scale-Tronix
4002-BT
NOTE
To safeguard during transport, the weighing unit inside
Babytherm has been protected against vibration and shock.
To ensure correct operation, this protection must be
removed before using the scale.
Before using the scale:
Loosen the locking screws (4x) under the bed by about
●
1 cm.
Before transporting Babytherm:
Tighten the locking screws (4x) under the bed.
●
Before each use
Check that the unit has been disinfected.
●
Check that the side panels are locked securely into place.
●
The panel hardware must be visible above the bed sur-
face.
Check that the side panels are free from cracks and sharp
●
edges.
Check that the correct mattress is in the cot;
●
Operation with mattress heater: gel mattress
Operation without mattress heater: foam mattress.
Check that the bed tilts properly and locks securely into
●
position.
Check that the required accessories and therapy equip-
●
ment are available and in proper working order.
Check that the gas supply is available and sufficient for the
●
accessories and equipment to be used.
Check that the cables and hoses are correctly and
●
securely installed. Never route cables or hoses over the
panels because they might be pinched or crushed when
folding up the panels or fitting the bed canopy.
Check whether both infrared radiators are providing
●
warmth by placing your hand carefully at a safe distance
from the radiant heater.