Recommended heater settings for heat
therapy
Instructions for setting the mattress heater and radiant
warmer, alone or in combination.
Using the mattress heater with a gel mattress without
using the radiant warmer (Babytherm 8010 only)
Normal setting:
The core temperature of the patient adapts to the mattress
temperature over a relatively long period. The mattress
heater should therefore be set to the appropriate core tem-
perature for the patient, e.g. between 38.0 °C and 38.3 °C for
a premature baby and 37.0 °C for a neonate born after a nor-
mal term. Monitor the core temperature of the patient and
adjust the mattress temperature setting to the needs of the
patient.
Always cover and/or clothe the patient. Do not place
●
blankets or other insulating material under the patient,
since they obstruct heat transfer to and from the bed
(warming/cooling).
Warm up the patient:
●
Set the mattress temperature setting to the desired core
temperature or, if necessary, slightly higher in order to
reduce the warming period.
Cool the patient:
●
Set the mattress temperature to a value lower than the cur-
rent core temperature, e.g. to 36.0 °C.