Read error memory
1 Press button until error mode is active.
2 The following message appears in the text display:
»Configuration Mode
Error memory«.
3 The error number appears: »FXX«.
4 Indication »Err«.
5 The frequency of the error is indicated.
P
p
6 Press
or
button and the next error is displayed.
7 Press
button. The equipment runs a self-test.
Check
G
8 Press
button for 3 seconds. The error memory is
erased.
Return to radiant warmer mode
Press OK button.