Operating Concept
Operating Concept
➀
ON/OFF indicator:
The green LED is lit when the unit is ON.
The red LED is lit following a power failure.
➁
OK button for confirming advisory messages and settings.
If the OK button is not pressed within 10 seconds of enter-
ing a new setting, the previous settings remain active.
➂
"Inop." alarm display.
➃
Pushbutton for checking the LEDs, displays, siren and
audible alarm.
➄
Pushbutton for switching off the audible alarm.
➅
Visual display for alarms:
Warning = red LED. Caution = yellow LED.
➆
Text display: shows advisory messages for the user.
➇
Babytherm 8010 only – display and keypad for the heated
gel mattress:
top: display for the measured mattress temperature;
bottom: display for the set mattress temperature,
with buttons for setting the desired temperature.
right: indicators for operation in extended setting
range – high (>38 °C) or low (<36 °C).
8
➈
°C
skin °C
manual
skin
man.
➉
➇
➆
Babytherm 8010
10
°C
5
> 38 °C
1
< 36 °C
mattress.°C
➈
Display and keypad for the radiant warmer:
top: display for the measured value of the core and periph-
eral skin temperature;
bottom: display for the set skin temperature, with buttons
for setting the desired value and a toggle button for select-
ing "man." (manual mode) or "skin" (baby control mode).
The active mode is indicated by the relevant yellow LED;
right: display of heater levels 1 to 10, for manual operation
of the radiant warmer:
Heater levels 1 to 3 = green LEDs
Heater levels 4 to 10 = yellow LEDs
complete with buttons for selecting the desired increment.
➉
On/Off buttons for:
Procedure light,
Night light,
Phototherapy, optional,
Radiant warmer.
Each of these buttons has a yellow LED that is lit when
the relevant function is switched on.
➅
➄
➃
Check
➂
Inop.
➁
OK
➀