Dräger Babytherm 8004 Instructions For Use Manual page 29

Open intensive care, resuscitation
Hide thumbs Also See for Babytherm 8004:
Table of Contents

Advertisement

Skin temperature above 39 °C
1 The skin temperature display starts flashing.
– The audible alarm sounds.
2 the red LED starts flashing.
– The red central alarm lamp on the radiant warmer starts
to flash.
3 The following message appears in the text display:
»Skin temperature above 39 °C
Press the OK button to acknowledge«.
2 The red LED lights.
The alarm can be muted for 2 minutes:
G
4 Press
button
or
5 Press the OK button.
– The audible alarm is cancelled.
2 The red LED is lit continuously.
– The red central alarm lamp on the radiant warmer goes
out.
3 The following message appears in the text display:
»Skin temperature above 39 °C«.
The radiant warmer should be switched off in manual
(»man.«) mode:
1 Skin temperature display is lit.
2 Red LED starts flashing.
The alarm is cancelled automatically when the measured skin
temperature drops below 38.5 °C again.
3
1
10
5
1
skin C
manual
mattress . C
skin
man .
Operation
Babytherm 8010
2
4
15 min
C
< 36 C
5
29