Care
Do not expose device to UV radiation as a means of
disinfecting/cleaning. Cracks may be caused in the acrylic
glass components.
Only use the recommended cleaning agents and disinfect-
ants. Otherwise, there is a danger of causing cracks in the
acrylic glass and macrolon, e.g. when using alcohol.
Use only preparations classified as "surface disinfectants" for
disinfecting.
For material compatibility, we recommend preparations
based on
– aldehydes,
– quaternary ammonium compounds.
Due to their chemical composition and the risk of damage to
materials, the following preparations are unsuitable:
– halogen-releasing compounds,
– strong organic acids,
– oxygen-releasing compounds.
The manufacturer's recommendation should always be fol-
lowed when choosing the product. The manufacturer is liable
for the data concerning suitable applications for the products
and any damage to property.
Do not use disinfectants or cleaning products that con-
tain alcohol.
For users in the Federal Republic of Germany, we recom-
mend the use of disinfectants listed in the current DGHM list
(DGHM = German Society for Hygiene and Microbiology).
The DGHM list (published by mhp-Verlag GmbH, Wiesbaden)
also specifies the active basis of each disinfectant. For coun-
tries where the DGHM list is unavailable, the above recom-
mendations apply. For example, the following wipe-
disinfectants may be used:
®
Dismozon
pur
Bode Chemie GmbH & Co, Hamburg
®
Incidur
Henkel Hygiene GmbH, Düsseldorf
®
Sekusept
powder Henkel Hygiene GmbH, Düsseldorf
Follow the Instructions for Use provided by the disinfectant
manufacturer.
NOTE
Even reusable accessories (e.g. after being prepared) have
a limited service life. Wear may be increased and the
service life reduced considerably by various factors when
handling and preparing them (e.g. disinfectant residues
corroding the material when autoclaving them). These parts
must be replaced if signs of wear become visible, such as
cracks, deformation, discoloration, peeling, etc.
54

Before next use

Reassemble components, pages 9 onwards.
After wipe-disinfecting, operate the fully assembled unit for
a few hours without a patient to eliminate any disinfectant
residues:
Set the mattress temperature to 37 °C
Set the radiant warmer to heat level 3 in »man.« mode.
Before a patient is next placed in the unit:
Fit all therapy accessories required.
Check that the unit is ready for operation, pages 18
onwards.