Dräger Babytherm 8004 Instructions For Use Manual page 27

Open intensive care, resuscitation
Hide thumbs Also See for Babytherm 8004:
Table of Contents

Advertisement

Deviations of a few tenths of a degree are normal.
NOTE
Therefore:
Only change the set value for the skin temperature if the
core temperature needs to be corrected.
Check, whether both infrared radiators are providing warmth,
by placing your hand carefully at a safe distance from the
radiant heater.
Outside measuring range
If the temperature is outside the measuring range of 15 °C
to 42 °C:
3 dashes at the bottom of the display = temperature
below 15 °C
3 dashes at the top of the display = temperature
above 42 °C
Refer to "Troubleshooting", see page 55.
If the temperature drops below 15 °C, the message
»Skin temp. below range« appears on the display.
The message »Skin temp. above range« appears if the
temperature rises above 42 °C.
If the sensor probe is disconnected or the sensor is
defective:
1 Three dashes light up in the middle of the display.
After 10 seconds:
– The audible alarm sounds,
2 the red LED flashes and
– the red central alarm lamp on the radiant warmer starts
flashing.
1 The three dashes on the display flash.
3 The display shows the message:
»Plug in skin temp. sensor
Press OK button to acknowledge«.
Or
»Skin temp. sensor fault
Press OK button to acknowledge«.
Immediately connect the sensor plug or change the skin
temperature sensor. The audible alarm can be muted
for 15 minutes:
G
4 Press
button
or
5 Press OK button.
– The audible alarm will be muted,
2 the red LED goes out and
– the red central alarm lamp on the radiant warmer goes out.
3 The display shows the message:
»Plug in skin temp. sensor«.
Immediately connect the sensor plug or change the skin tem-
perature sensor.
If the error cannot be remedied immediately:
6 Switch to manual mode (»man.«), see page 23.
°C
°C
3
1
10
5
1
skin C
manual
mattress . C
6
skin
man .
Operation
Babytherm 8010
2
4
15 min
C
< 36 C
5
27

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Babytherm 8010

Table of Contents