Operation And Setting Of The Menu; Function; Calibration - sauter KERN SW 1000 Instruction Manual

Sound level meter
Hide thumbs Also See for KERN SW 1000:
Table of Contents

Advertisement

4. operation and setting of the menu

4.1 Function

4.2 Calibration

calibration factor. For calibration with the sound calibrator, the method By
SW-BA-e-2020
Octave table page 3.
display of the measured
data of LAeqLBeq, LCeq
and
LZeq.
lights red and
when the data exceeds
the threshold value.
Press the <Menu> button to enter the next menu. All measurement
parameters can be set in the menu.
Structure menu:
Select the function and press the < Enter> button to enter this
menu. You can select 2 measurement types: Level measurement
and 1/1 octave measurement. Press the arrow keys<▲>, <▼> to
select the measurement mode. Press the <Enter> button to save
the setting and return to the previous menu. Press the <ESC>
button to return to the previous menu.
Select the Calibration option and press the < Enter> button to enter
this menu.
The sensitivity of the microphone is influenced by numerous factors,
such as temperature, humidity and air pressure. Therefore, the user
must perform at least one calibration prior to measurement.
There are two calibration procedures. By measurement and With
The
LED
appears
19