Maintenance and care

Maintenance to be performed by the
customer
For the protection of persons, environment and mate-
rial you are obliged to clean the centrifuge regularly
and to disinfect it if necessary.
Unsuitable cleaning agents or disin-
fection procedures may damage the
centrifuge and its accessories!
Before using cleaning agents or dis-
infection
mended by the manufacturer, the
user have to make sure by consulting
the manufacturer, that the procedure
foreseen does not cause any dam-
ages to the instrument!
Instruments with refrigeration unit:
If a strong ice sheet is present in the internal
chamber, be sure to remove all condensate
after defrosting!
procedures
not
recom-

Cleaning

Pull mains plug before cleaning the in-
strument!
Clean the casing, the rotor chamber, the rotor and the
accessories regularly and in case of need. This is indi-
cated both for reasons of hygiene and to prevent cor-
rosion due to contamination sticking to the instrument
and its accessories.
Clean them with mild agents of pH values ranging from
6 to 8.
Immediately after cleaning, dry the aluminum parts or
put them into a warm-air dryer at a temperature not
exceeding 50°C.
During cleaning liquids and espe-
cially organic solvents should not
come into contact with the drive
shaft and the ball bearing.
Organic solvents may decompose
the lubricant of the motor bearing
causing the drive shaft to lock.