Control panel
Display panels
Speed / RCF
Resting state:
current value (0) or pre-selected set-point value
During run:
present current value of speed or RCF
(after activation of switch key)
Circulating light-points:
rotor turns
End:
„End"
Error codes:
flashing display with error code
Run time
Resting/End:
current value (0) / "End", or pre-selected set
point value
(in minutes, or "hd" for permanent operation)
Time selection/run: - remaining time till "0" in minutes
Permanent operation (hd):
„Quick run":
- run time passed
Temperature*
according to:
current sample temperature in °C
(in temperature equilibrium)
* only for instruments with refrigeration unit
- run time passed in seconds/ minutes
in seconds/ minutes
Keys
start:
start of the centrifuge
stop:
manually stop of a run
open lid:
open lid
(possible only with the instrument switched on and
standstill of the rotor)
quick run:
short-term run up of the centrifuge as long as key
remains pressed
Switch
Speed/RCF:
switching between speed and RCF display
Pre-temp:
pre-temp-function*
"set" keys:
stepwise increase/decrease of set-point values
Short pressing of any of the "set" keys:
switch from current to pre-selected set-point value, signaled by
flashing display.
(Error codes (see „troubleshooting"section):
E-14: over temperature centrifuge chamber (> 50 °C)
E-22: troubled speed measurement
E-24: lid will not open
E-31: over temperature motor
E-36: over amperage or over voltage
E-46: lid was opened manually during run
E-57: imbalance
E-60: under temperature centrifuge chamber (< -20 °C)