Fisher Scientific Isotemp 148 Instruction Manual
Fisher Scientific Isotemp 148 Instruction Manual

Fisher Scientific Isotemp 148 Instruction Manual

General purpose and flammable materials storage refrigeration

Advertisement

Quick Links

Instruction Manual
â
Fisher Scientific Isotemp
General Purpose
and
Flammable Materials Storage
Refrigeration
GP Freezer Model 148, 148HK
FMS Freezer Model 149, 149HK
FMS Refrigerator Model 152, 152HK

Advertisement

Table of Contents
loading

Summary of Contents for Fisher Scientific Isotemp 148

  • Page 1 Instruction Manual â Fisher Scientific Isotemp General Purpose Flammable Materials Storage Refrigeration GP Freezer Model 148, 148HK FMS Freezer Model 149, 149HK FMS Refrigerator Model 152, 152HK...
  • Page 2 Service Division (Repairs): 1-800-395-5442 Technical Support: 1-800-926-0505 Customer Service: 1-800-766-7000 Prior to calling for service, have the following information readily available: Catalog Number: Serial Number: Date of Purchase: The catalog number and serial number can be found on the data plate located on the left interior wall of the unit.
  • Page 3: Table Of Contents

    Table of Contents Isotemp GP and FMS Refrigertors Introduction............................1 Unpacking............................1 Visible Loss or Damage.........................1 Concealed Loss or Damage ......................2 Installation............................2 Selecting a Location........................2 Leveling The Unit ..........................2 Shelves ............................3 Electrical Connection........................3 Operation ............................4 Initial Startup ..........................4 Temperature Control........................4 Condensate Disposal........................4 Material Compatibility........................5 Trouble Shooting Table........................6 Maintenance .............................7...
  • Page 4: Introduction

    Page Isotempâ GP & FMS Refrigeration Introduction Your satisfaction and safety are important to Fisher Scientific and a complete understanding of this unit is necessary to attain these objectives. As the ultimate user of this apparatus, it is your responsibility to understand its proper function and operational characteristics.
  • Page 5: Concealed Loss Or Damage

    2 of 8 Page Isotempâ GP & FMS Refrigeration Concealed Loss or Damage Concealed loss or damage refers to loss or damage, which does not become apparent until the merchandise has been unpacked and inspected. Should either occur, make a written request for the carrier’s agent within 15 days of the delivery date;...
  • Page 6: Shelves

    3 of 8 Page Isotempâ GP & FMS Refrigeration Shelves Shipped inside each cabinet are four fixed shelves which also serve as part of the evaporator coil. Determine the total amount of current presently being used by Electrical Connection other apparatus connected to the circuit that will be used by this refrigerator.
  • Page 7: Operation

    4 of 8 Page Isotempâ GP & FMS Refrigeration WARNING: DO NOT under any circumstances cut or remove the third (ground) prong from the power cord. DO NOT use a two-prong adapter plug. Where a two-prong wall receptacle is encountered, it is the personal responsibility and obligation of the user to have it replaced with a properly grounded three-prong receptacle.
  • Page 8: Material Compatibility

    5 of 8 Page Isotempâ GP & FMS Refrigeration Material Compatibility The interior cabinet of this unit is constructed of epoxy- coated steel. Care must be exercised when determining which chemicals may be stored in the refrigerator and freezer sections and which type of storage materials should be employed.
  • Page 9: Trouble Shooting Table

    Trouble Shooting Table This table is intended to assist in resolving user-correctable Refrigerator problems by relating symptoms to their likely causes. If service beyond the scope of this table is required, contact your nearest Fisher Scientific Service Office. Symptom Probable Cause Action...
  • Page 10: Maintenance

    7 of 8 Page Isotempâ GP & FMS Refrigeration Maintenance CAUTION: When servicing the unit, disconnect from the electrical power source Cabinet Cleaning The cabinet interior should be cleaned frequently. Any spilled liquid should be wiped off immediately. Stains resulting from some spills can be permanent if not quickly removed.
  • Page 11: Performance Characteristics

    8 of 8 Page Isotempâ GP & FMS Refrigeration Performance Characteristics Temperature Ranges 151: 3° to 13°C (37° to 55°F) 152: 3° to 13°C (37° to 55°F) 148: -20° to –10°C (-4° to 14°F) 149: -20° to –10°C (-4° to 14°F) Temperature Stability @ 4 °C ±4 °C @ -20 °C ±4 °C...
  • Page 12 Part No. 104180 Rev C Fisher Scientific Co. Isotemp GP & FMS Refrigeration 2000 Park Lane Published 6-03 Pittsburgh, PA 15275-112 Technical Service: 1-800-943-2006...