Rotors For The Accuspin Micro 17R - Fisher Scientific AccuSpin Micro 17 Instructions For Use Manual

Microcentrifuges
Table of Contents

Advertisement

Accessories

Rotors for the accuSpin Micro 17R

Table 3: Performance characteristics for accuSpin Micro17R
rotor designation
order no.
places / volume
maximum permissible load [ g ]
-1
minimum speed n
[ min
min
-1
maximum speed n
[ min
max
maximum RCF value at n
max
acceleration / deceleration time [ s ]
radius maximum / minimum [ cm ]
angle
[ ° ]
minimum temperature at n
max
relative to room temperature 23°C
aerosol-tight *
permissible temperature range
autoclavable (number of cycles)
* Tested and approved by HPA, Porton-Down, UK – Please see the references in the chapter "Aerosoltight application"!
14
Microliter-rotor
24 x 2 ml
75003524
24 x 1,5 / 2 ml
24 x 4
]
300
]
13 300
17 000
10 / 12
8,6 / 5,1
45
[ °C ]
< 0
yes
-4 °C to +40 °C
121°C, (20 cycles)
differences of 120V instruments are shown in parentheses
Microliter-rotor
36 x 0,5 ml
75003436
36 x 0,5 ml
36 x 0,5
300
13 300
15 600
8 / 10
7,9 / 5,0
45
< 0
no
-4 °C to +40 °C
121°C, (20 cycles)
Dual-rotor
18 x 2 ml / 0,5 ml
75003418
18 x 2 ml + 18 x 0,5 ml
18 x 4 + 18 x 0,5
300
13 300
16 800
10 / 12
8,5 / 4,8
45
< 0
no
-4 °C to +40 °C
121°C, (20 cycles)

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Accuspin micro 17r

Table of Contents