Operation

Pretemp

The pretemp-function allows quick and easy tempering
of the unloaded rotor.
After calling the function by pressing the key
the desired temperature still must be chosed.
(If any other key is pressed, the pretemp function will
be quit).
After activating the „start" key
ated at optimal speed until the the desired tempera-
turen is reached.
An LED above the "
activated pretemp function.
32
the rotor will be oper-
" key indicates operation at the
If you wish to change the temperature of your samples,
please consider that the time required for temperature
adjustment is prolonged. The farther apart initial and
only
final temperature, the longer it takes for the tempera-
ture to adjust.
The temperature reading does not give the
change in the temperature of the sample
(the temperature display reading is delayed
with respect to the /actual/ sample tempera-
ture change). You can neither witness the
heating nor the cooling of the samples di-
rectly. For critical applications you should
take other precautions to ensure that the
desired temperature is actually reached and
maintained (e.g. by measuring the tempera-
ture immediately after the centrifugation run).