Maintenance and care
5. Treat the rotor and the rotor lid according to the
instructions given for the disinfectant (soaking in
liquid or spraying). You must strictly observe the
specified action times!
6. Turn the rotor head down and drain off the disinfec-
tant. Thereafter thoroughly rinse rotor and lid with
water.
7. Dispose of the disinfectant according to valid regu-
lations.
8. Aluminum rotors have to be treated with anticorro-
sive protective oil subsequently.
9. All seals have to be lubricated again.
40
Disinfection with bleaching lye
These agents contain highly ag-
gressive hypochlorites and must
not be used with aluminum rotors!
The following precautionary measures are to be taken
for extensive protection of the plastic rotors:
1. Avoid high temperatures!
The bleaching solution and the rotor should not be
warmer than ca. 25 °C.
2. Do not let the bleaching solution act longer than
absolutely necessary!
3. After disinfection, rinse the rotor thoroughly with
distilled water and allow drying.
4. All seals have to be lubricated again