Troubleshooting
Error
Display "oP" appears
although lid is closed.
"Lid"
46
Symptom
Exceptionally running
noise.
Will not start.
Rotor stops with
deceleration to standstill.
Possible causes and corrective measures
Imbalance.
1. Stop the centrifuge by pressing the "stop" key, in case of
emergency, unplug mains power cord.
2. Wait until the centrifuge comes to a complete stop.
3. Check whether the rotor is properly loaded.
4. Check whether a broken tube, damage to the rotor or
motor is responsible for the run noise.
If you cannot locate and solve the problem yourself, contact a
service representative
Centrifuge lid not properly closed
- Open the lid and repeat locking procedure.
If you cannot locate and solve the problem yourself, contact a
service representative.
Centrifuge lid was opened manually during the run.
- Close centrifuge lid immediately
The instrument phases out with deceleration.
For further centrifugation, you have to switch the instrument
off and switch it on again.