Removing the rotor

To remove the rotor you have to proceed in the reserve
order as at the installation of the rotor.
In case of contamination you can remove the rotor from
the drive shaft without opening the aerosol lid. Then
you can open the removed rotor e.g. in a safety work
bench and decontaminate it.
1. Open the centrifuge lid.
2. Unscrew the rotor nut with the provided. Turn the
rotor nut counterclockwise.
3. Grab the rotor centrically and carefully pull it per-
pendicularly off the drive shaft. Make sure not to tilt
it.

Audible alarm

Accompanying all error messages, a warning signal is
given out which can be silenced upon pressing any
key.
By default there is an acoustic signal at the end of any
centrifugation run. You have the possibility to switch off
this signal. For this, you have to press the „speed/ RCF
switch key"
while switching on the centrifuge.
Depending on the pre-selected mode the display
shows the following signs: "Snd" "on" or „Snd" „oF".
By pressing the set keys
acoustic signal function can be turned off or on. Sub-
sequently, by pressing the stop key
will be activated.
Operation
or
in the field "time" the
the new setting
35