Maintenance and care

Maintenance to be performed by the
customer
For the protection of persons, environment and mate-
rial you are obliged to clean the centrifuge regularly
and to disinfect it if necessary.
Unsuitable cleaning agents or disin-
fection procedures may damage the
centrifuge and its accessories!
If you intend to use cleaning agents
or disinfection procedures not rec-
ommended by the manufacturer, you
have to make sure by consulting the
manufacturer,
foreseen does not cause any dam-
ages to the instrument!
that the procedure

Cleaning

Pull mains plug before cleaning the
instrument!
Clean the casing, the rotor chamber, the rotor and the
accessories regularly and in case of need. This is indi-
cated both for reasons of hygiene and to prevent cor-
rosion due to contamination sticking to the instrument
and its accessories.
Clean them with mild agents of pH values ranging from
6 to 8.
For other cleaning agents please consult Fisher -
Scientific
!
Immediately after cleaning, dry the aluminum parts or
put them into a warm-air dryer at a temperature not
exceeding 50°C.
Maintenance and care
47