Factory Default Settings - GE OptiSonde User Manual

General eastern chilled mirror hygrometer
Table of Contents

Advertisement

October 2007

Factory Default Settings

Function
Analog Output A
Humidity, Units: Tdew °C, Range: –40°C to +60°C
Analog Output B
Temperature, Units: Tmp °C, Range: 0 to 100°C
Alarm
Disabled, Humidity, Units: Tdew °C, Set Point: Upper: 0.000, Lower:0.000
Auto Balance
Interval: Daily at 12:00 noon, Auto Cycle, Enabled.
Data Field 1
Humidity, Units: Tdew °C
Data Field 2
Temperature, Units: Tmp °C (Dry bulb)
Sound
On
Gas
Molecular Weight of Gas: 28.9645 (typical for air)
Communication
Baud rate: 38,400
Parameters
Humidity, Units: Tdew °C
Serial Outputs
Humidity, Units: % RH
Temperature, Units: Tmp °C
Status: Stopped, Interval: 60 second, Separator: tab, Parameters: Td °C,
Data Log [COMM]
%RH; Flags: enabled*
Status: None/Stopped, Interval: 60 second, Separator: tab,
Data Log [FILE]
Parameters: Td °C,
Pressure
101.325 kPa (typical atmospheric pressure @ sea level).
*A status flag gives an indication, via the RS-232 interface, of the OptiSonde status, such as
"Control" and "PACER," as well as the state of the alarm ("ALARM" or "_______").
3-4
As shipped from the factory, the OptiSonde is normally programmed
with the configuration shown in Table 3-2 below.
Table 3-2: Factory Default Values
Ta°C, %RH; Flags: enabled*
Setting
Ta°C,
Operation