5. Communication

Note: Default settings in the parameter tables are indicated with an asterisk (*)
unless explicitly stated.

5.1. Device Address

Sets the device address or poll ID on a HART network. Any address other than 0 will
cause the output current to be a fixed value, and the current will not indicate the
reading.
Values
To reset Device Address via SIMATIC PDM:
Open the project in Process Device Network View then right-click on the
device.
Go to Object Properties > Connection to access the field Short Address.

6. Security

Note: Default settings in the parameter tables are indicated with an asterisk (*)
unless explicitly stated.

6.1. Remote Access

6.1.1. Access Control
Note: If access control is changed to limit remote access, it can only be reset
via the handheld programmer.
Enables /disables the read/write access to parameters via remote
communications.
Options
7ML19985JP02
SITRANS LR200 (HART) – INSTRUCTION MANUAL
Range:
0 to 15
Default: 0
No changes are permitted via remote communica-
Read only
tions.
*
Read Write
Changes are permitted.
Sets the status to Read only, with the potential for
Restricted
another HART device to change this parameter
only, via remote communications.