Tvt Shaper - Siemens Sitrans LR200 Instruction Manual

Hart
Hide thumbs Also See for Sitrans LR200:
Table of Contents

Advertisement

TVT Shaper

Note: Double click on each axis to see the Xscale and Data Scale values. Right-click or
Left-click on the axis and drag to reposition the scale.
This feature allows you to manually adjust the TVT to avoid false echoes caused by
obstructions. (For an explanation see
Open the menu Device – Echo Profile Utilities and click on the tab TVT Shaper
Click on Measure to refresh the echo profile and load the current TVT from the
device.
Change the position of the cursor on the TVT using the Point+ and Point– buttons:
raise and lower the TVT using Offset+ and Offset–.
Alternatively, enter values for Point and Offset directly into the dialog boxes.
Click on Transfer to Device.
7ML19985JP02
Auto False Echo Suppression (2.8.7. 1 .)
cursor
TVT
SITRANS LR200 (HART) – INSTRUCTION MANUAL
on page 126.)
echo
profile
Page 47