Linearization
You can use the linearization feature to define a more complex vessel shape and enter up
to 32 level breakpoints where the corresponding volume is known. The values
corresponding to 100% and 0% levels must be entered. The breakpoints can be ordered
from top to bottom, or the reverse.
Example (values for example purposes only)
Level
Breakpoint
value
number
4
3
2
1
Using Linearization via the Quick Start wizard
Open the menu Device – Wizard - Quick Start:
1)
In Step 1 – Identification, click on Read Data from Device, select language, and click
on Next.
2)
In Step 2 – Application, select a volume application, for example Volume in a vessel,
and click on Next.
3)
In Step 3 – Vessel Shape, choose the vessel shape option Linearization Table, and
click on Next.
7ML19985JP02
Break-
Level
point
value
no.
(m)
19.5
1
0
2
5
9
3
9
5
4
19.5
0
SITRANS LR200 (HART) – INSTRUCTION MANUAL
Volume
0.5 m
value
(l)
0
500
20 m
3000
8000
Page 41