Configuring A Stillpipe Application - Siemens Sitrans LR200 Instruction Manual

Hart
Hide thumbs Also See for Sitrans LR200:
Table of Contents

Advertisement

5)
In Step 5 – Summary, check parameter values. Click on BACK to return and revise
values, FINISH to save settings offline, or FINISH AND DOWNLOAD to save settings
offline and transfer them to the device.
The message Quick Start was successful will appear. Click on OK.

Configuring a stillpipe application

Launch the Quick Start wizard and follow steps 1 to 2.
In step 2, select Application Type – Level in a vessel with stillpipe, Material – Liquid
Low dK, and set the stillpipe diameter as desired.
The wizard updates the propagation factor according to the pipe diameter, and
enables the CLEF algorithm for low dK liquids (see
page 123 for more detail).
Continue through steps 3 to 5 then click on FINISH AND DOWNLOAD to save
settings offline and transfer them to the device.
7ML19985JP02
SITRANS LR200 (HART) – INSTRUCTION MANUAL
Position Detect (2.8.4.2.)
on
Page 43