Operating Mode; Low Calibration Point; High Calibration Point; Apply? (Apply Changes) - Siemens Sitrans LR200 Instruction Manual

Hart
Hide thumbs Also See for Sitrans LR200:
Table of Contents

Advertisement

1.5. Operating Mode

1.6. Low Calibration Point

1.7. High Calibration Point

1.8. Apply? (Apply changes)

2. Setup

Note: Default settings in the parameter tables are indicated with an asterisk (*) unless
explicitly stated.

2.1. Device

2.1.1. Hardware Revision
Corresponds to the electronics hardware of the SITRANS LR200.
2.1.2. Firmware Revision
Corresponds to the firmware that is embedded in the SITRANS LR200.
2.1.3. Loader Revision
Corresponds to the software used to update the SITRANS LR200.
2.1.4. Order Option
Read only. Displays the device type: standard or NAMUR NE 43-compliant.

2.2. Sensor

2.2.1. Units
Sensor measurement units.
Values
2.2.2. Sensor Mode
sensor
value
Mode
NO
SERVICE
LEVEL
SPACE
DISTANCE
a)
See Material Level (2.5.1.) for more detail.
7ML19985JP02
SITRANS LR200 (HART) – INSTRUCTION MANUAL
m, cm, mm, ft, in
Default: m
Distance
High
Calibration
Space
Point
Level
Far Range
Description
Measurement and associated loop
current are not updated, and the
device defaults to Fail-safe mode
*
Distance to material surface
sensor reference point
Dimensions: Uni-
(see
Construction Polypropylene Rod
Antenna
on page 13 and
Appendix G: Flange Adapter
Versions
on page 150)
Low
Calibration
Point
Reference point
n/a
a)
.
Low Calibration Point
(process empty level)
High Calibration Point
(process full level)
Sensor Reference Point
Page 75