Specifications; Power; Performance - Siemens Sitrans LR200 Instruction Manual

Hart
Hide thumbs Also See for Sitrans LR200:
Table of Contents

Advertisement

Specifications

Siemens Milltronics makes every attempt to ensure the accuracy of these
Note:
specifications but reserves the right to change them at any time.

Power

General Purpose:
Non-incendive (FM/US only):
Intrinsically Safe:
Flameproof:
Increased Safety:
Explosion-proof (FM/CSA US/Canada only):
• Maximum 30 V DC
• 4 to 20 mA
• Max. startup current

Performance

Reference operating conditions according to IEC 60770-1
• ambient temperature
• humidity
• ambient pressure
Measurement Accuracy (measured in accordance with IEC 60770-1)
- From end of antenna to 600 mm (1.96 ft)
- Remainder of Range
Frequency
Max. measurement range
Min. detectable distance
• 3", 4", and 6" horn
• 8" horn
• PP rod, 100 mm internal shield
• PP rod, 250 mm internal shield
• PTFE rod, unshielded
• PTFE rod, 100 mm external shield
• PTFE rod, 250 mm external shield
1)
From sensor reference point. For the sensor reference point for each configura-
Dimensions: Uni-Construction Polypropylene Rod Antenna
tion, see
for the standard version, or
other configurations.
2)
3" and 4" horns should be used only in stillpipe applications.
Page 8
SITRANS LR200 (HART) – INSTRUCTION MANUAL
Startup Behavior
see
1)
1)
2)
Appendix G: Flange Adapter Versions
Nominal 24 V DC at
max. 550 Ohm
Nominal 24 V DC at
max. 250 Ohm
on page 138
+15 to +25 °C (+59 to +77 °F)
45% to 75% relative humidity
860 to 1060 mbar g (86000 to 106000 N/m
40 mm (1.57")
10 mm (0.4") or 0. 1 % of span (whichever is
greater)
5.8 GHz (6.3 GHz in N. America): refer to
device nameplate for confirmation
20 m (65.6 ft)
300 mm (11.8")
330 mm (12.9")
417 mm (16.4")
567 mm (22.3")
417 mm (16.4")
474 mm (18.6")
624 mm (24.5")
on page 150 for
2
g)
on page 13
7ML19985JP02