Select Analog Output; Self Test - Siemens Sitrans LR200 Instruction Manual

Hart
Hide thumbs Also See for Sitrans LR200:
Table of Contents

Advertisement

2)
Modify desired values, and click on Write.
3)
Click on Read, to see the effects of your modification.
4)
If desired, set reminders for either or both of Reminder 1 (Required)/Reminder 2
(Demanded).
5)
Click on Snooze to add a year to the Total Expected Device Life.
To set Service/Calibration schedules:
1)
Open the menu Device – Maintenance, and click on the Service/Calibration
Schedule tab.
2)
Modify desired values, and click on Write.
3)
Click on Read, to see the effects of your modification.
4)
If desired, set reminders for either or both of Reminder 1 (Required)/Reminder 2
(Demanded).
5)
Click on Service/Calibration Performed to reset the schedule.

Select Analog Output

Allows you to set the mA Output to report Level, Distance, or Space (see
on page 80 for an illustration).
Function (2.6. 1 .)
If a volume application is selected, mA Output is automatically set to Volume. (See Analog
Output on page 128 for more details.)
1)
Open the menu Device – Select Analog Output.
2)
The Select Analog Output window displays the current setting: click on OK.
3)
Select a different setting and click on OK.
4)
The Select Analog Output window displays the new setting: click on OK.

Self Test

Checks memory (RAM and Flash). If there are no errors, returns the message 'Self Test
OK.' If errors are found, returns the message 'Self Test Fails' .
Open the menu Device – Self Test, select the option Yes and click on OK.
7ML19985JP02
SITRANS LR200 (HART) – INSTRUCTION MANUAL