4.5.3. Time until Next Calibration
Read only. Calibration Interval (4.5.1.) less Time since Last Calibration (4.5.2.).
4.5.4. Reminder Activation
Note: To modify this parameter via SIMATIC PDM it must be accessed via the
pull-down menu Device – Maintenance.
Allows you to enable a maintenance reminder.
Values
a) First set the values in Reminder 1 (Required) (4.3.5.)/Reminder 2
(Demanded) (4.5.6.).
b) Select the desired Reminder Activation option.
4.5.5. Reminder 1 (Required)
If Time until Next Calibration (4.5.3.) is equal to or less than this value, the device
generates a Maintenance Required reminder.
Values
a) Modify values as required.
b) Set Reminder Activation (4.5.4.) to the desired option.
4.5.6. Reminder 2 (Demanded)
If Time until Next Calibration (4.5.3.) is equal to or less than this value, the device
generates a Maintenance Demanded reminder.
Values
a) Modify values as required.
b) Set Reminder Activation (4.5.4.) to the desired option.
Page 106
SITRANS LR200 (HART) – INSTRUCTION MANUAL
* Timer OFF
ON - no reminders checked
ON - Reminder 1 (Maintenance Required) checked
ON - Reminders 1 and 2 checked
ON - Reminder 2 (Maintenance Demanded) checked
Range: 0 to 20 years
Default: 0.164 years
Range:
0 to 20 years
Default: 0.019 years