4.3.8. Acknowledge Status
Indicates which level of maintenance reminder has been acknowledged.
In SIMATIC PDM, open the menu View – Device Status, click on the
Maintenance tab and check the Sensor Lifetime Status window.
4.3.9. Acknowledge
Acknowledges the current maintenance reminder.
To acknowledge a reminder via SIMATIC PDM:
a) Open the menu View – Device Status and click on the tab Maintenance.
b) In the Device Lifetime section, click on Acknowledge Warnings.
To acknowledge a reminder via the handheld programmer:
a) Press RIGHT arrow
Mode.
b) Press RIGHT arrow

4.4. Service Schedule

Notes:
• Four sets of parameters allow you to monitor the Device/Sensor Lifetimes and
set up Maintenance/Service schedules, based on operating hours instead of a
calendar-based schedule. See also Remaining Device Lifetime (4.2.), Remaining
Sensor Lifetime (4.3.), and Calibration Schedule (4.5.).
• Performing a reset to Factory Defaults will reset all the Maintenance Schedule
parameters to their factory defaults.
• The device operates in years. To view Service Interval parameters in hours or
days (via SIMATIC PDM only) see Service Interval (4.4.1.).
The device tracks service intervals based on operating hours and monitors the
predicted lifetime to the next service. You can modify the Total Service Interval, set
schedules for maintenance reminders, and acknowledge them.
The maintenance warnings and reminders are communicated to the end user
through status information. This information can be integrated into any Asset
Management system. For optimal use, we recommend that you use SIMATIC PCS7
Asset Management Software in conjunction with SIMATIC PDM.
To access these parameters via SIMATIC PDM:
Open the menu Device – Maintenance and select the Service Schedule tab.
After modifying values/units as required, click on Write to accept the change,
and Read to view the effect of the change.
Click on Service Performed to restart the timer and clear any fault messages.
7ML19985JP02
SITRANS LR200 (HART) – INSTRUCTION MANUAL
twice to open parameter view and activate Edit
to acknowledge the reminder.
(see illustration on next page)
Page 101