Sliding Waveguide on Anaerobic Digesters

Notes:
• Minimum distance from the flange face to the target is limited by Near Range (2.8. 1 .).
• We recommend adjusting Auto False Echo Suppression parameters with an empty
or almost empty vessel.
This application is to obtain a level measurement and corresponding 4 to 20 mA output
proportional to the sludge level on a digester.
sensor
reference
point
High
Cal.
Point
10 m
Low
Cal.
Point
Parameter
type
Quick start
wizard
parameters
Independent
parameters
Page 142
SITRANS LR200 (HART) – INSTRUCTION MANUAL
2 m
horn
Parameter Name (number)
Material (1.2.)
Response Rate (1.3.)
Units (1.4.)
Operating Mode (1.5.)
Low Calibration Point (1.6.)
High Calibration Point (1.7.)
Apply? (Apply changes) (1.8.)
Timer (2.5.2.)
Material Level (2.5.1.)
• The raised position is for installation
and maintenance.
• The lowered position is for operation.
• Program the instrument for operation
in the lowered position.
• Low Cal. Pt. is the bottom of the
digester (10 m/33 ft from sensor
reference point when the instrument
is lowered to its normal operating
position).
• High Cal.Pt. is 2 m (6.56 ft) from sensor
reference point.
• The maximum rate of filling or
emptying is about 0. 1 m (4")/min.
Value
Function
LIQUID
MED
1 m/min. (3.28 ft/min.)
M
meters
LEVEL
Level
10
10 m/33 ft)
2 m (6.56 ft
2
YES
2
2 minutes
HI
Hi
)