Sirona C2+ Cleaning And Care page 19

Hide thumbs Also See for C2+:
Table of Contents

Advertisement

Sirona Dental Systems GmbH
Care and cleaning instructions for the practice team C2
A
SELECT ITEMS
SELECT ITEMS
FOR PURGING
FOR PURGING
0.2
1
1
pm x 1000
59 58 512 D 3370
D 3370.103.01.13.02
+
+
, M1
wrong
right
0.2
1
rpm x 1000
6.2 Purge/Auto Purge function (purging the water paths)
On pressing quick setting key 1 briefly (< 3 s), the fol-
lowing prompt appears if all instruments are deposited in
their holders:
SELECT ITEMS FOR PURGING
Remove the instruments to be purged from the dentist
and assistant elements and insert them in the adapters
of the sanitation bowl as far as they will go.
ATTENTION
Be careful not to snap off the instrument hoses when in-
serting the instruments.
The maximum water flow must be set on the instru-
ments.
Press quick setting key 1 again (< 3 s), the instruments
are then purged one after the other.
The purging time in seconds set previously in the Mode
dialog runs down for each instrument individually.
An audible signal sounds after the last instrument has
been flushed.
You now can continue in either of the following ways:
Leave the instruments in the sanitation bowl:
The instruments will automatically be purged as
soon as the treatment center is switched on
again. The treatment center can then be
switched off again.
The APURGE function remains activated.
The instruments are returned to their holders:
The treatment center can be prepared for daily
practice operation.
The APURGE function is completed.
i
NOTE
When removing the sanitation bowl, please notice that it
is filled with water.
Canceling the APURGE function
1
The purging process can be canceled as follows
• Press intensity key 1 for > 3 seconds
or
• Place the instruments in their holders
6 Disinfection system