Cuspidor - Sirona C2+ Cleaning And Care

Hide thumbs Also See for C2+:
Table of Contents

Advertisement

2 Cuspidor

2
Cuspidor
5
6
Amalgam
6
Grease
Care and cleaning instructions for the practice team C2
After each patient
Clean the cuspidor.
ATTENTION
Use only cleaning agents which have been approved by
Sirona (see page 2, General information)!
ATTENTION
Do not use any domestic cleaning and care agents.
Foaming domestic cleaning agents lead to foam and wa-
ter being drawn into the dry suction system, resulting in
damage to the suction machine.
Daily during the noon break and at the end of
each workday
Remove the gold trap (5) and clean it.
Store the amalgam residues under water in a separate,
closed container. When replacing the amalgam rotor
(see page 12), put the residues into the rotor for dis-
posal.
After the cleaning process, fill at least 1 liter of water into
the cuspidor for flushing.
ATTENTION
Do not dispose of amalgam residues caught in the gold
trap in the cuspidor or the sink!
Weekly
Pull out the tumbler outlet (6) or the hydrocolloid
adapter, if applicable, and clean it.
Before refitting the tumbler outlet or the hydrocolloid
adapter, lightly lubricate the O-rings.
ATTENTION
Use only lubricants which have been approved by Sirona
(see page 2, General information)!
Sirona Dental Systems GmbH
+
+
, M1
59 58 512 D 3370
D 3370.103.01.13.02